All IMMULITE 2000 Automated Immunoassay Analyzers with SoftwareVersion 1.1, a clinical automated immunoassay analyzer intended
 for in-vitro diagnostic use for detection of analytates in human
 specimens. Recall #Z-909-9.
REASON
 If the initial counts of the reaction tubes are exactly 10,000
 cps, this leads to erroneously low calculated results.
CODE
 All serial numbers in commercial distribution (Note: The recall
 product was initially released with Software Version 1.1 on March
 2, 1998).
MANUFACTURER
 DPC Cirrus, Inc., a subsidiary of Diagnostic Products
 Corporation, Randolph, New Jersey.
RECALLED BY
 Manufacturer, by telephone March 2 & 3, 1999, and by Technical
 Bulletins dated March 2 and 3, 1999. Firm-initiated field
 correction complete.
DISTRIBUTION
Nationwide and international.
QUANTITY
334 units were distributed.