Venturi Ventilator, provides a software that continuously adaptsto the patient’s changing breathing demands and assures ventilation at the lowest pressure. Recall #Z-803-9. REASON Software error may cause ventilator to deliver more than set tidal volume. CODE Venturi Ventilator with Software Revision C. Serial #970222. MANUFACTURER Cardiopulmonary Corporation, Milford, Connecticut. RECALLED BY Manufacturer, by letter on...