Mennen Medical Ltd. 1/21/98 Patient Monitors c. The approval page for the Horizon 9000WS Cathlab’s software located in the document control department contained the signature of the originator of the document, but it lacked an approval signature. This approval page was different from the approval page for the same software located in the production department....