PIE Medical Equipment Netherlands ultrasound

PIE Medical Equipment 4/1/96 Netherlands ultrasound

1. Failure to maintain records documenting that production and quality assurance measurement equipment is routinely calibrated, inspected, and checked according to written procedures and to validate software programs by adequate and documented testing, when computers are used as part of an automated production or quality assurance system, as required by 21 CFR 820.61. For example:

a) Manufacturing, assembly and test operations are controlled by paperless routings [purged text]. There was no documentation of the entry and testing of each medical device routing to ensure that the data entered, procedural references, data fields and test specification error codes are in accordance with the Device Master Record.

b) There were no requirement specifications or test documentation for the RS232 serial integration with either the Scanner 200 diagnostic ultrasound system or the Bio-Tek 601 Pro automated safety tester.

c) There were no requirement specifications that describe the intended functions of the complaint, Unix or DOS database systems in terms of data fields, field sizes, field types, screens, sorts, system security, data integrity, data backups and user locations. Testing has not been conducted to assure and verify that the database system performs reliably and consistently.

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.