Recall – Physiomonitoring System IBP values freeze on display

Company: Invivo Corporation
Date of Enforcement Report: 11/14/2018
Class II:

PRODUCT
Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Recall Number: Z-0462-2019

REASON
The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

RECALLING FIRM/MANUFACTURER
Invivo Corporation, Orlando, FL on 9/24/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
76

DISTRIBUTION
Distributed to the Netherlands

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
Registration Link

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.