Cl ll HORIBA ABX PENTRA 400 I.S.E. Module

Company: Horiba Instruments, Inc dba Horiba Medical .
Date of Enforcement Report: 4/24/2013
Class ll:

PRODUCT

HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes. Recall Number Z-1141-2013

REASON
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.

RECALLING FIRM/MANUFACTURER
Horiba Instruments, Inc dba Horiba Medical, Irvine, CA on 3/20/2013. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
905 units.

DISTRIBUTION
Nationwide.